Intelligent intracranial thrombectomy
Summary
On March 17, 2026, Sensome (France) and Robocath jointly completed the world’s first proof-of-concept trial of intracranial thrombectomy using a smart thrombus-sensing guidewire combined with a vascular interventional robot. In a silicone brain model with an implanted animal thrombus, the study verified that the Clotild smart guidewire is fully compatible with Robocath’s commercial R-One system and its next-generation neurointerventional robot under development, achieving in-situ identification and precise localization of thrombus components under robotic manipulation. This study is the first to demonstrate the feasibility of integrating biosensing and robotic control technologies, and provides technical support for the implementation of remote neurointerventional surgery.
I. The Clotild® Smart Guidewire System
The Clotild Smart Guidewire System, developed by Sensome, a French company, is a research-based minimally invasive intelligent guidewire system. It is the world’s first thrombus characterization device integrating impedance microsensing technology, and has not yet received global commercial approval. This product addresses the clinical challenge of unknown thrombus composition during mechanical thrombectomy. Through miniature sensors at the guidewire tip, it simultaneously performs in-situ biological characterization of the thrombus during navigation, providing operators with tissue biological information that traditional imaging methods cannot obtain. The product was developed and manufactured in collaboration with ASAHI INTECC, and its core sensing technology has been validated in multi-center clinical studies.
1.1 Working Principle
Clotild’s core technology is based on multi-frequency electrochemical impedance spectroscopy (EIS), which measures the electrophysiological properties of fluids or tissues within a 360° radius around the sensor. Analyzing this information using its proprietary machine learning predictive model, it can identify fresh thrombi rich in red blood cells and other types of thrombi in real time. While tissue impedance measurement is widely used in the diagnosis and treatment of easily accessible lesions, such as tumor diagnosis and atrial fibrillation ablation, it has never been used for thrombus detection due to the large size of existing technologies. Sensome has successfully miniaturized the technology and integrated it into the distal portion of a standard 0.014” guidewire, located behind the soft, non-invasive guidewire tip, creating the world’s smallest impedance sensor. The system interface provides the surgeon with a clear assessment of the thrombus’s nature during the procedure.
Its core technological principles can be summarized in the following three points:
1. Tissue Impedance Acquisition: The microelectrode at the guidewire tip emits multi-frequency weak electrical signals to the surrounding tissue. Different biological tissues (erythrocyte-rich thrombi, fibrin-rich thrombi, blood vessel walls, and normal blood) exhibit unique electrical impedance spectrum characteristics due to differences in cell density, membrane permeability, and extracellular matrix composition.
2. Real-time Signal Analysis: The acquired impedance spectrum data is transmitted in real-time to an external processing terminal via the guidewire’s built-in circuitry. Sensome’s proprietary machine learning model, trained on over 1000 clinical thrombus samples, performs real-time classification and quantification of tissue impedance.
3. Bioinformation Output: The model can output core biological characteristics of the thrombus, such as erythrocyte percentage, fibrin percentage, and tissue stiffness, within one second. Simultaneously, it accurately distinguishes whether the guidewire tip is contacting a thrombus, blood vessel wall, or normal blood, providing real-time data for the operator’s decision-making.
1.2 Structural Composition
The core structure of the Clotild system follows the design principles of minimally invasive compatibility and functional integration, integrating four major functional modules without changing the traditional 0.014-inch guidewire outer diameter:
1. **Advanced Microsensor Unit:** Integrates the world’s smallest impedance sensing electrode, measuring only a few hundred micrometers, encapsulated at the distal end of the guidewire to ensure guidewire passability and vascular compliance.
2. **Flexible Guidewire Body:** Utilizes a medical-grade nickel-titanium alloy substrate with built-in micro-signal transmission circuitry, balancing guidewire pushability, torsional control, and signal transmission stability, compatible with all conventional interventional procedures.
3. **Proximal Signal Interface:** A standardized physical connection interface that seamlessly interfaces with external signal processing terminals, enabling lossless transmission of sensor data, while also being compatible with guidewire control modules in robotic systems.
4. **AI Algorithm Terminal:** An external signal processing and analysis unit responsible for impedance data amplification, filtering, and AI model inference, outputting real-time visualized thrombus biological information.
The Clotild guidewire’s functional structure is similar to conventional interventional guidewires, with a diameter of 0.014 inches and a length of 2 meters, adaptable to commercially available microcatheters. The access establishment process is identical to that of conventional guidewires.
1.3 Existing Clinical Evidence

Background: In endovascular mechanical thrombectomy, recanalization of intracranial vascular occlusion fails in approximately 20% of cases, and thrombus composition is considered a significant factor affecting treatment efficacy. This study aimed to demonstrate that combining electrical impedance spectroscopy (EIS) with machine learning can serve as a novel and highly accurate method for identifying different types of thrombi in humans.
Methods: This study included 134 samples, categorized into four groups for EIS analysis: 29 “white thrombi,” 26 “mixed thrombi,” 12 “red thrombi,” and 67 liquid “blood” samples. Thrombi were prepared in vitro from citrate-anticoagulated human blood from five healthy volunteers. Histological analysis was used to classify and validate thrombi based on red blood cell count. A machine learning prediction model was constructed based on impedance data to distinguish blood from various thrombus samples and further differentiate between the four different sample categories.
Results: Histological analysis confirmed that the composition of in vitro prepared thrombi was similar to that of clinically obtained human thrombi. The predictive model achieved a sensitivity of 90% and a specificity of 99% in distinguishing between blood and thrombi; the overall accuracy in classifying the four sample types was 88%.
Conclusion: Combining impedance spectroscopy with machine learning provides an efficient method for differentiating between different types of thrombi and liquid blood. These results suggest the potential development of a probe-like device (such as a neurovascular guidewire) with an impedance-based sensor integrated at its distal end, enabling real-time characterization of thrombus properties upon contact. This information could help operators develop better thrombectomy strategies, thereby improving clinical outcomes for stroke patients.

Background: In the process of mechanical thrombectomy (MT) for acute stroke, real-time characterization of thrombi during the procedure could guide instrument selection and improve recanalization rates. Electrochemical impedance spectroscopy (EIS) has been used for real-time characterization of various biological tissues, but its application in thrombosis research has not been fully realized. Objective: To conduct a feasibility study using EIS analysis on thrombus samples obtained through mechanical thrombectomy to evaluate:
(1) the ability of EIS combined with machine learning to predict the red blood cell (RBC) content in thrombi;
(2) the accuracy of classifying thrombi as “RBC-rich” or “RBC-poor” at different RBC content thresholds.
Methods: The ClotbasePilot study was a multicenter, international, prospective feasibility study. Histological analysis was performed on thrombi obtained through thrombectomy to determine the proportions of RBCs and other components. Analysis was performed using EIS data combined with machine learning methods, and linear regression was employed to assess the correlation between EIS results and histological analysis. The sensitivity and specificity of the model in classifying thrombi as RBC-rich or RBC-poor were also evaluated.
Results: In 514 cases of mechanical thrombectomy, 179 thrombus samples were included for EIS and histological analysis. The mean RBC content in the thrombi was 36% ± 24%. A good correlation was found between EIS predictions and histological analysis (regression slope 0.9, R² = 0.53, Pearson correlation coefficient = 0.72). At different RBC thresholds (20%–60%), the model’s sensitivity for thrombus classification was 77%–85%, and its specificity was 72%–88%.
Conclusion: EIS combined with machine learning can reliably predict the RBC composition of isolated acute ischemic stroke thrombi and classify them based on their composition, demonstrating good sensitivity and specificity. This method provides a potential technical pathway for intraoperative thrombus characteristic assessment and individualized thrombectomy strategy development.
II. Robocath’s Robotic Systems
Robocath, a leading French company specializing in vascular interventional robotics, was founded in 2009. It focuses on developing biomimetic master-slave vascular interventional robotic systems to address pain points in traditional interventional procedures, such as operator radiation exposure, insufficient operational precision, and difficulties in implementing telemedicine. Its core commercial product, R-One, is the first coronary interventional robotic system in Europe to receive CE certification and has also been approved for market launch in China by the NMPA. It has been used in over 500 clinical procedures at more than 30 centers worldwide. Currently, the company is developing a next-generation neurovascular interventional robotic system, optimizing micro-manipulation capabilities for the tortuous anatomy of cerebral blood vessels, aiming to expand the application of robotics in the field of neurointervention.
2.1 Working Principle
The Robocath robotic system is based on master-slave bionic manipulation technology. Its core principle is to precisely replicate the surgeon’s hand movements into high-precision mechanical motions: The surgeon, using a bionic handle at the master console, performs operations such as pushing, rotating, and positioning the guidewire/catheter. The handle accurately captures fine movements like twisting and pushing. The central control system processes the operation signals in real time, filtering out minute hand tremors and compensating for motion based on vascular anatomy to ensure operational stability. The slave mechanical unit, installed beside the operating table, precisely replicates the surgeon’s movements based on the processed signals, achieving sub-millimeter-level guidewire/catheter pushing and continuous 0-360° rotation. It also features a built-in force feedback module that senses the contact force between the instrument and the blood vessel wall in real time, preventing vascular damage.
This technology isolates the surgeon from radiation sources while improving operational precision to the micrometer level, addressing the precision and radiation concerns of traditional manual operations.
2.2 Structural Composition
The R-One system comprises a remotely controlled robotic unit and a mobile radiation protection control workstation. The remotely controlled robotic unit includes the robot body and an adjustable support robotic arm. Disposable sterile operating components include catheter access, spare access, guidewire access, guidewire clamping button, and robotic arm release brake. The mobile radiation protection control workstation includes a control unit and a mobile radiation protection screen. This workstation is equipped with a display, catheter control joystick, guidewire control joystick, fluoroscopy pedal, angiography control module, and operating table and C-arm control system.

2.3 Existing Clinical Evidence
Background: Robot-assisted angiography represents a promising but nascent technology in endovascular interventional therapy.
Objective: To report the first human experience with robot-assisted angiography in extracranial carotid circulation.
Methods: Patients with extracranial carotid circulation lesions who visited Houston Methodist Hospital between March and September 2019 were included in this study.
Results: Six patients met the inclusion criteria. Five underwent only diagnostic angiography with robot-assisted catheterization, while one received delayed therapeutic intervention after diagnosis. The mean age was 51 ± 17.5 years. The mean anesthesia time was 158.7 ± 37.9 minutes, the mean fluoroscopy time was 22.0 ± 7.3 minutes, and the mean radiation dose was 815.0 ± 517.0 mGy. No technical complications or postoperative neurological deficits occurred in any case. No cases required conversion to artificial neurovascular intervention.
Conclusion: Introducing robotic technology into neurovascular intervention (NVI) can help improve surgical proficiency and reduce occupational exposure risks. Although the safety and effectiveness of this technology have been validated in the coronary and peripheral vascular fields, sufficient evidence is still lacking in the neurovascular field. This study reports for the first time the feasibility of robot-assisted neurovascular intervention in extracranial carotid circulation in humans.
III. Combined Application of Clotild® and Robocath Robotic Systems
This joint research endows the robot with tissue-level tactile perception through Clotild’s biosensing capabilities. While verifying the compatibility of intelligent sensing guidewires with vascular interventional robot systems and providing a reference for the general integration of intelligent interventional devices and robots, it also accumulates multimodal data of tissue biological information and robot operation data through real-time sensing data acquisition under robot operation. Ultimately, by leveraging the robot’s remote control capabilities and the experience-free perception of intelligent guidewires, it breaks through geographical limitations, promotes the downward flow of high-quality neurointerventional medical resources to remote areas, and realizes the popularization of thrombectomy.
This joint proof-of-concept was conducted in a silicone cerebral circulation model implanted with a porcine thrombus. Using Robocath’s commercial R-One system and its next-generation neurointerventional robot (currently under development), coupled with the Clotild smart guidewire, a simulated intracranial thrombectomy was performed, achieving three core validation results:
1. System-wide compatibility verification: The Clotild smart guidewire is fully compatible with the Robocath robot’s guidewire control module. The robot can accurately navigate the guidewire within tortuous cerebral blood vessels, with lossless transmission of the guidewire’s sensor signals;
2. Precise thrombus characterization under robotic manipulation: During robot-controlled guidewire manipulation, Clotild successfully achieved in-situ real-time identification of thrombi in the model, accurately distinguishing between the thrombus and the vessel wall, and quantifying the thrombus’s component characteristics, providing real-time tissue perception information for robot operation;
3. Enhanced operational safety: Compared to the traditional guidewire + robot combination, the addition of Clotild allows the robot to perceive the tissue type at the guidewire tip in real time, effectively avoiding the risk of the guidewire accidentally contacting the vessel wall, reducing trial-and-error steps, and significantly shortening the surgical operation time.
IV. Some Thoughts
The joint research between Sensome and Robocath represents a fusion of biological and operational intelligence, providing an intelligent pathway for neurointerventional thrombectomy. In the future, with the integration of more intelligent technologies such as intracranial OCT catheters, magnetically controlled guidewire robots, and AI image analysis platforms, clinicians may be truly freed from the physically demanding and experience-based tasks of monitoring images and inhaling radiation, focusing instead on the entire process and making final decisions. For large medical device manufacturers and distributors, the focus may shift from simply selling consumables and equipment to selling complete intelligent surgical solutions and data services. Smaller and medium-sized enterprises will need to develop compatible products by integrating with the real-time ecosystem.
In terms of R&D capabilities, companies not only need to establish an advantage in a specific core technology but also possess interdisciplinary integration capabilities, developing a systematic understanding and collaborative ability in areas such as active sensing technology, robotic systems, artificial intelligence, and medical imaging.