Pulmonary embolism + combination of drugs and medical devices + broad market

Thrombolex’s core product is the Bashir Endovascular Catheter (BEC), a pharmacomechanical thrombolysis device specifically designed for acute pulmonary embolism (PE).

The BEC employs a pharmacomechanical thrombolysis strategy, combining targeted thrombolytic drug delivery with mechanical thrombus fragmentation. A nitinol-reinforced spiral cutting basket at the catheter tip repeatedly expands and retracts within the thrombus, creating multi-planar fissures for physical thrombus fragmentation.

The basket has a maximum diameter of 45 mm when fully expanded, and its length is reduced by approximately 2.5 cm when fully expanded. The significantly increased surface area of ​​the fragmented thrombus provides more fibrin binding sites for both endogenous and exogenous fibrinolytic drugs; simultaneously, the central channel formed by the basket expansion promotes blood flow restoration within the thrombus, accelerating the thrombolysis process. Diluted tissue plasminogen activator (tPA) is pulsed into the thrombus through 48 laser-drilled holes, forming drug bubbles trapped within the thrombus, continuing to exert thrombolytic effects even after catheter withdrawal.

The BEC is a 7F catheter system. Its main components include a spiral-cut infusion basket consisting of six mini infusion catheters with 48 laser-drilled holes, available in both long (12.5 cm) and short (10 cm, Bashir Short-Basket) sizes to accommodate pulmonary arteries of varying lengths. The basket is reinforced with a nickel-titanium alloy, providing excellent radial support and repositionable deployment.

The delivery system is compatible with 0.018-inch guidewires, with a total length of 92.5 cm, and can be inserted via a 7F intravenous sheath. A deployment actuator is located at the handle, allowing the operator to control the expansion and contraction of the basket.

The standard version (Bashir) features a single-arm drug delivery channel, while the Plus version (Bashir Plus) upgrades to a dual-arm design for more efficient drug infusion. Drugs can be administered via syringe or infusion pump using pulsatile spray or continuous infusion. Size compatibility: Suitable for vessels with diameters of 3–45 mm, covering treatment needs from the main pulmonary artery to segmental pulmonary arteries.

The product has been approved by the FDA 510(k) and its safety and efficacy have been validated in the RESCUE multicenter prospective study, showing a 33.3% reduction in the right ventricle/left ventricle ratio and a 35.9% reduction in pulmonary artery obstruction within 48 hours, with a low incidence of bleeding complications and device-related adverse events.

Pulmonary embolism