Monitoring of end-tidal carbon dioxide
Carbon dioxide measurement is performed during the ventilation cycle, monitoring the expiratory carbon dioxide concentration at the time of airway opening and the corresponding time or expiratory volume, and then plotting the carbon dioxide concentration graphically. The peak value (peak value) of the expiratory carbon dioxide graph represents the end-tidal carbon dioxide level (PETCO2), which is equivalent to the partial pressure of carbon dioxide in arterial blood (PaCO2). PETCO2 is typically 1–5 mmHg lower than PaCO2, ranging from 35–43 mmHg.
Because normal PETCO2 is very close to PaCO2, it is often used as a non-invasive monitoring device for ventilation efficiency. Abnormal waveforms in PETCO2, such as large fluctuations in carbon dioxide levels or low PETCO2, may indicate altered ventilation and perfusion distribution or airway obstruction. A larger difference between PaCO2 and PETCO2 [P(a-ET)CO2] may be due to positive pressure ventilation therapy (especially with positive end-expiratory pressure (PEEP)), pulmonary embolism, cardiac arrest, or hypoperfusion of the lungs.
1. Principle
(1) Chemical Method: The disposable calorimeter is a colorimeter based on a chemical reaction, providing a qualitative estimate of exhaled carbon dioxide concentration. When carbon dioxide is below 3 mmHg, the colorimeter is purple; if carbon dioxide is between 3% and 15%, it is yellowish-brown; if carbon dioxide is above 15 mmHg, the colorimeter will turn yellow due to carbon dioxide acidification.
(2) Spectroscopic Method: Provides quantitative data on exhaled carbon dioxide concentration.
1) Infrared Spectroscopy: Carbon dioxide absorbs infrared light at a maximum wavelength of 4.26 μm. As the concentration of exhaled carbon dioxide increases, the amount of infrared light absorbed also increases, which is reflected in the monitor.
2) Mass Spectroscopy: Exhaled gas is collected in a vacuum chamber, where it is ionized by electron beams. The gas content is then determined based on mass and charge. 3) Raman spectroscopy: When ultraviolet or visible light strikes gas molecules, energy is absorbed, and new, longer wavelengths are emitted, forming a red light shift spectrum that reflects the content of exhaled gas.
4) Gas sampling monitoring methods: Common clinical methods include sidestream and mainstream. Sidestream analyzers can be used for intubated (endotracheal tube) or non-intubated patients, with sampling rates of 150 cc/min and 50 cc/min. The sampling time and the time for data presentation often differ. Mainstream analyzers place the sensor directly on the artificial airway to directly monitor and analyze the patient’s exhaled carbon dioxide.
2. Physiological and Technical Issues
(1) Cardiopulmonary diseases: Since the carbon dioxide value of exhaled breath is primarily measured at the partial pressure of carbon dioxide at the end of exhalation, individuals with cardiopulmonary diseases will have a lower PETCO2 than PaCO2 due to increased dead space.
(2) The presence of water or N2O can interfere with carbon dioxide readings because the wavelengths at which CO2 absorbs infrared light are close to those of water and N2O. Therefore, the presence of water or N2O in the gas being measured will interfere with the carbon dioxide reading.
(3) Because the carbon dioxide content in exhaled breath increases with the duration of exhalation, it is necessary to determine the timing of the monitor reading and the patient’s condition beforehand.
3. Clinical Applications
(1) Indications
1) Monitoring the severity of lung disease and assessing treatment response, particularly focusing on information related to improved body volume or ventilation/perfusion.
2) Confirming correct endotracheal tube placement.
3) Monitoring the correct placement of the artificial airway in mechanically ventilated patients during treatment.
4) Monitoring the ventilation efficiency in mechanically ventilated patients during treatment.
5) Monitoring the efficiency of pulmonary and coronary blood flow.
6) Monitoring body carbon dioxide production.
(2) Contraindications: No absolute contraindications.
(3) Hazards/Complications
1) Mainstream monitors: ① May increase airway ineffective volume; caution should be exercised with treatments that are too small; ② Increased weight increases the chance of airway slippage in pediatric patients.
2) Lateralflow monitors: ① Flow-driven ventilators may increase the risk of automatic ventilation; ② Reduced volume. (4) Monitoring: When monitoring carbon dioxide levels or concentrations, the following parameters must be considered simultaneously:
1) Tidal volume, respiratory rate, positive end-expiratory pressure (PEEP), inspiratory/expiratory ratio (I/E ratio), peak airway pressure, and therapeutic oxygen concentration.
2) Hemodynamic parameters, such as blood pressure, pulmonary pressure, cardiac output, shunts, and ventilation-perfusion ratio.
4. Graphic Interpretation
The value of exhaled carbon dioxide is related to the patient’s respiratory cycle. The waveform of exhaled carbon dioxide concentration over time can be divided into four stages.
(1) Stage 1: Primarily the volume of the dead space in the airway, therefore the CO2 concentration is very low.
(2) Second stage: Alveolar CO2 is mixed with gas in the dead space (which does not contain CO2). Therefore, the CO2 concentration increases as the dead space gas decreases and the alveolar gas volume increases. Thus, we can see that the CO2 vs. time graph gradually increases over time.
(3) Third stage: This stage is close to the end of expiration. CO2 continues to flow out of the alveoli, reaching its peak. The change in CO2 over time shows a plateau-like phenomenon, known as the plateau phase.
(4) Fourth stage: In this stage, the patient has begun inhalation, so the CO2 content begins to decrease rapidly, returning to zero.
If the end-expiratory CO2 curve continues to rise without a plateau, it indicates that the expiratory time is too short. If the CO2 curve does not return to zero at the start of the next inhalation, it indicates CO2 re-inhalation. At this time, the maximum exhaled CO2 value will also gradually increase.
If the exhaled CO2 is 0 or drops to 0 after 5 breaths, it indicates that the endotracheal tube may not be in the trachea.