New Explorations in Below-Knee Vascular Interventional Therapy: Application of OCT and FFR(II)
(IV) Differences in Vessel Diameter and Stenosis Assessment
1. Diameter Difference: In 40% (4 cases), the target healthy vessel lumen diameter estimated by the physician based on DSA differed from the OCT measurement by more than 0.5 mm.
2. Stenosis Difference: In 60% (6 cases), the stenosis degree estimated by the physician based on DSA differed from the OCT measurement by more than 25%.
DISCUSS
(I) Core Findings
This study is the first proof-of-principle study to apply OCT and FFR in 10 patients with below-knee PTA, confirming their safety, feasibility, and lack of complications. In clinical application, the combination of these two methods can alter intraoperative decisions in 70% of patients, primarily in the following ways:
1. Avoiding overtreatment: Some patients, assessed by FFR, showed no hemodynamic significance in their lesions and required no treatment. However, treatment of proximal lesions in the lower leg may not resolve ischemic pain and tissue loss caused by severe microcirculatory impairment in the foot, and may even induce dissections and thrombosis not visible on DSA but visible on OCT, increasing the risk of early re-occlusion. This may explain the low patency rate of lower leg PTA.
2. Optimizing treatment plans: In some patients, postoperative FFR showed hemodynamic significance in the lesions, while OCT revealed dissections or thrombosis, requiring further treatment (such as balloon dilation or stent placement). However, not all dissections are clinically significant, and further research is needed to clarify which postoperative dissections require further management.
(II) Relevance to Previous Studies
1. OCT-Related Studies: The results of this study are consistent with previous studies on the application of OCT in below-knee PTA. A case report by Marmagkiolis et al. showed that OCT was used in popliteal artery PTA in patients with limb ischemia without perioperative complications, and OCT detected intraluminal protrusion of calcified plaques not shown by DSA. The results after additional angioplasty were ideal. A study by Paraskevopoulos et al. showed that OCT successfully imaged 19 out of 21 stents in assessing in-stent stenosis in the lower leg without serious perioperative complications, confirming its safety and feasibility.
2. FFR-related studies: The results of FFR altering intraoperative decision-making in 6 patients in this study are similar to those of Ruzsa et al. (whose study confirmed the feasibility and safety of FFR in 39 patients with chronic limb-threatening ischemia). However, Ruzsa et al.’s study lacked descriptions of perioperative complications and relevant data on the impact of FFR on surgical decisions. In addition, the FFR threshold (0.8) used in this study is derived from cardiology guidelines, and its applicability in the calf artery needs to be verified by subsequent studies.
(III) Limitations of the Study
1. Small Sample Size: Only 10 patients were included, making it impossible to analyze in detail the impact of OCT and FFR on intraoperative decisions. Larger clinical trials are needed to validate their value in below-knee PTA.
2. Cost Considerations: OCT and FFR measurements incur additional costs. Larger-scale studies are needed to verify whether the benefits of reducing overtreatment outweigh the costs.
3. Uncertain FFR Threshold: The effectiveness of currently used cardiology FFR thresholds in the lower leg artery is unclear. Further research is needed to determine suitable FFR thresholds for the lower leg artery.
CONCLUSION
Optical coherence tomography (OCT) and fractional flow reserve (FFR) are safe and feasible for use in percutaneous endovascular angioplasty (PTA) below the knee. Both techniques provide information on the hemodynamic significance, size, and morphology of lesions that was previously unavailable, significantly altering perioperative treatment decisions and offering a new direction for precision treatment of below-knee arterial lesions. However, further studies with larger sample sizes are needed to validate their long-term clinical value, determine suitable FFR thresholds for lower leg arteries, and assess cost-effectiveness.
References:
van Wely, C., Oosterveld, R.J., Bouwman, L.H. et al. Optical coherence tomography and fractional flow reserve in below-the-knee percutaneous transluminal angioplasty: a pilot study. CVIR Endovasc 8, 87 (2025). https://doi.org/10.1186/s42155-025-00580-9