CT technology and neurointervention

Background:

The first clinical brain CT scan was performed in 1972 by Hounsfield using an improved prototype CT scanner. The first CT scanner, the EMI Mark 1, was used clinically in 1973. In recognition of Godfrey Hounsfield and Alan Cormack's contributions to computed tomography and their impact on clinical medicine, the 1979 Nobel Prize in Medicine was awarded to them. CT scanning technology has since been continuously improved. Slip ring and spiral scanning techniques were first proposed by Varian and further refined by Toshiba and Siemens in 1985 and 1987, respectively. An EMI Mark I scan takes approximately 5 minutes to complete one slice, while a spiral CT scan can cover several centimeters within 60 seconds.
Multi-slice CT scanning was first realized on the EMI Mark I first-generation scanner, capable of acquiring images of two layers simultaneously. In 1992, Elscint introduced the first spiral CT scanner utilizing dual detectors; from 1998 onwards, spiral CT evolved to four-slice or more. Now, 320-slice CT scanners are available, capable of covering the entire head (16cm) in a single scan, achieving near-isotropic resolution and 4D imaging. Using multi-slice CT, both thin and thick slice images can be effectively acquired simultaneously when covering a considerable distance along the Z-axis. The acquired thin-slice image data can be combined with reconstructed thick-slice images for image interpretation, preserving the advantages of thin-slice scanning (such as minimizing volumetric and linear artifacts). Given the original data, multi-planar, three-dimensional, and four-dimensional images can be reconstructed from thin-slice images.

CT technology

Traditional CT and CT Spinal Imaging/CT Cistern Imaging:
Head CT images can be obtained using continuous single-slice, spiral multi-slice, or multi-slice techniques. The American College of Radiology guidelines recommend that the scan thickness for supratentorial lesions should be ≤10mm in adults and ≤5mm in children under 10 years of age; for posterior fossa lesions, the thickness should be ≤5mm in both adults and children. For skull base lesions, thin-slice plain scans of ≤3mm are preferred. If multiplanar reconstruction (e.g., fractures of the craniofacial region after trauma) or 3D reconstruction is required, the slice thickness must be ≤2mm. Using multi-slice CT scans with a slice interval of ≤1mm allows for the simultaneous acquisition of multiple data points. For example, while performing a standard head CT, data from the face, skull base, or temporal bone can be acquired. Thin-slice images can be used to assess the brain and soft tissues, and the face, skull base, or temporal bone can be evaluated by combining thin-slice images with reconstructed images. The same applies to spinal CT after myelography or head CT after ventriculography. Coronal CT can acquire data by adjusting the machine angle and the patient's head position. The new MDCT system can acquire high-quality reconstructed data from axial images; however, if the reconstructed data cannot provide sufficient detail for clinical use, direct coronal scans are still required. Furthermore, if the patient moves during the scan, the quality of the reconstructed images will degrade.
For trauma and neurological emergencies, including ischemia and intracranial hemorrhage, CT is the preferred imaging modality. For most of the brain parenchyma, skull base, skull top, and spine, plain CT scans are usually sufficient. Intravenous contrast-enhanced CT is primarily used for inflammation and tumors, but its practical application is limited because central nervous system inflammation and tumors are typically diagnosed with MRI, thus eliminating the need for contrast-enhanced CT. Contrast-enhanced CT is generally used for soft tissue lesions in the neck and for CTA. For a penetrating injury, a plain CT scan of the neck can assess the extent of the injury, the condition of surrounding blood vessels, and the presence of foreign objects such as bullets or fragments. Myelography and ventriculography inherently require contrast enhancement.

CT angiography (CTA)

For rapid assessment of neck and intracranial vessels, CTA has become a superior alternative to digital subtraction angiography (DSA). A complete CT scan includes plain scan, head or neck vascular imaging, and CT perfusion imaging, with the total time for contrast-enhanced imaging generally being 10–15 minutes. CTA is well-tolerated and, in many ways, equivalent to MRI and MRA, making it readily accepted by patients. Raw images can be acquired immediately and provide most of the diagnostic information for clinical diagnosis. Data from the head and neck can be acquired simultaneously in a single scan, allowing for reconstruction of the Circle of Willis with smaller slice intervals (e.g., 0.5 mm) and a smaller field of view (FOV) using the raw data. Volumetric reconstruction, multiplanar reconstruction, and maximum intensity projection (MIP) images can be rapidly performed on a console or in a standalone workstation.
The typical method is to intravenously inject 70 to 100 ml of iodinated contrast agent at a rate of 3 to 4 ml/s. During this process, continuous thin-section scanning of the axial position is performed, and then a small amount of normal saline is added. The 50% overlap of the images during reconstruction is beneficial to improving the resolution in volume reconstruction and reconstructed images. The bolus injection is timed according to the filling status of the contrast agent in the aortic arch and left ventricle, which allows optimal arterial visualization and avoids venous interference. There are two timing methods, one is to use professional software to automatically trigger timing, and the other is to use the contrast agent bolus fraction to obtain enhanced data during continuous data collection. For the display of the circle of Willis, the standard delay time is about 25 seconds, and the arteries can be well visualized. Of course, some adjustments are required for patients with poor cardiac function. CT venography (CTV) only needs to be delayed by 6 to 8 seconds on the basis of the above time. The specific parameters vary from machine to machine and need to be optimized.
There are many image post-processing techniques, the most commonly used including Volume Reconstruction (VR), Motion Injection (MIP), and reconstruction of inclined/frontal curved surfaces. MIP is a ray tracing algorithm that calculates and projects the brightest pixels along each ray on a scanned object onto a plane, and is very useful. Volume Reconstruction distinguishes blood vessels from surrounding bone and soft tissue based on differences in color and transparency; it is helpful for observing intracranial lesions, especially aneurysms. VR technology has relatively limited use for neck imaging because of overlapping neck structures, such as the difficulty in processing veins and the spinal cord, but in certain cases, such as carotid endarterectomy, it is helpful in distinguishing high-positioned carotid bifurcations from the mandible. Surface Masking (SSD) has limited effectiveness and often underestimates the degree of vascular stenosis; it has now been replaced by other reconstruction techniques. Other post-processing techniques include automated vascular analysis (calculating the severity of stenosis associated with vascular calcification and extracting vascular images), software for separating arteries and veins, and "flyover" techniques. Bone and metal subtraction or masking algorithms can better visualize skull base aneurysms and are used for aneurysm detection and treatment planning. The density contrast between blood vessels and adjacent bone or calcifications can overlap, which limits the application of threshold segmentation methods; dual-energy bone subtraction can be used in this case. Regardless of the presence or absence of subtraction, related artifacts such as hard wire artifacts can affect diagnostic accuracy, a significant problem when routinely using aneurysm clips, metal coils, and stents. Overall, CTA is well-suited for evaluating stents, while MRI is better suited for evaluating coils; compared to MRA, CTA is more successful at visualizing the proximal end of aneurysm clips, and some aneurysm clips, such as titanium alloy aneurysm clips, are visualized even better.

CT perfusion

In clinical practice, two main imaging techniques are used to measure cerebral blood flow (CBF): (1) intravascular or non-diffusion tracing (bolus contrast agent) techniques; and (2) diffusion tracing techniques (where the tracer diffuses outside the blood vessels). Most CT and MRI perfusion-enhanced images utilize intravascular contrast agents, with the contrast agent injected intravenously to assess its initial passage through brain tissue. Diffusion tracing techniques include xenon CT perfusion imaging, PET-CT imaging, and arterial spin labeling MRI perfusion imaging; Wintermark and Latchaw et al. have discussed the differences between these techniques in their articles.

Mass contrast contrast agent CT perfusion

Contrast-enhanced CT perfusion (CTP) imaging is based on the linear relationship between iodine concentration and density in brain tissue. One to four slices are scanned in cine mode, acquiring one image every 1 to 2 seconds before, during, and after contrast agent injection. These slices are scanned repeatedly for a total of 40 to 50 seconds. A total volume of 40 to 50 ml of contrast agent is injected via a large vein at a rate of 5 to 8 ml/s. A series of CT scans are analyzed to obtain time-density curves (TACs). Patient movement significantly affects the analysis of perfusion-enhanced images. Although image analysis software can reduce errors to some extent by reconstructing these motion-affected data, the patient must be kept as still as possible before the examination. Sedation may be used if necessary.
Hemodynamic parameters include time to peak contrast (TTP, the time from the arrival of the contrast agent in the tissue to the peak level in the tissue, in seconds), cerebral blood volume (CBV, the total blood volume in 100g of brain tissue, including capillaries and large vessels, in ml/100g), mean transit time (MTT, the average time for the contrast agent to travel from an artery to a vein, in seconds), and cerebral blood flow (CBF, the blood flow through 100g of brain tissue per unit time, in ml/(100g·min)). TTP and CBV can be calculated directly and are relatively easy to obtain, while MTT and CBF calculations require more complex methods, including measuring the arterial input function (deconvolution technique) or assessing the morphology of the contrast-enhanced angiography (TAC) (non-deconvolution technique). Deconvolution techniques can correct for incomplete contrast agent boluses (theoretically assumed to be instantaneous boluses, but in reality, they are continuously injected), removing imperfections in the arterial input function relationship from the TAC, preserving the remaining function, from which CBF and MTT are obtained. The central volume law can explain the relationship between these parameters, such as CBF = CBV/MTT.
Non-deconvolution techniques utilize the slope of the contrast-enhanced angiography (TAC) to measure changes in blood iodine concentration over time, which is directly proportional to brain fluid flow (CBF) and the concentration difference of iodine in arterial and venous blood. The contrast agent injection rate is 6–8 ml/s, unlike deconvolution techniques, which can have an injection rate <5 ml/s. Deconvolution techniques can quantitatively analyze absolute CBF values, but their accuracy is difficult to guarantee; for example, they may overestimate CBF values ​​when analyzing large vessels in the region, while non-deconvolution techniques may underestimate CBF values. Clinically, relative values ​​of perfusion parameters are often used for analysis, with normal brain tissue or contralateral brain tissue as internal controls. Another method, different from deconvolution and non-deconvolution techniques, is a percentage perfusion volume map generated by subtracting the original CTA image from the baseline CT image. In this map, the density change of brain parenchyma is directly proportional to the iodine solubility in the tissue.

Xenon CT

Stable xenon perfusion CT can be used to measure absolute CBF (cerebrovascular flow). Xenon is lipid-soluble and easily diffuses; changes in its density can be measured on CT. Xenon CT perfusion imaging can be applied to many clinical conditions such as cerebrovascular diseases, traumatic brain injury, balloon occlusion tests, subarachnoid hemorrhage, and cerebral vasospasm. However, the FDA has not yet approved its clinical use, primarily because xenon has been reported to cause sleep apnea and increased intracranial pressure; these adverse reactions are transient and technically manageable, especially when using low concentrations of xenon (28%), whereas the original concentration was 33%.
First, baseline images were obtained by scanning 2–8 layers twice without xenon inhalation; then, these layers were scanned 6 more times during xenon inhalation. To measure changes in xenon density, the average density of each layer was calculated by subtracting the density of those layers at subsequent time points. A 28% xenon-oxygen mixture was inhaled, and the end-tidal xenon concentration was measured, along with monitoring end-tidal carbon dioxide solubility and any apnea. The CBF value was calculated using Fick's law, which states that the amount of tracer in a tissue sample is proportional to the difference between arterial inflow and venous outflow. A modified Kety-Schmidt equation was used to describe the relationship between the blood-brain partition coefficient and CBF in the distribution of xenon in the brain and arterial blood. Xenon solubility in brain tissue was obtained by measuring baseline and CT values ​​during xenon inhalation. Time-dependent arterial blood xenon solubility could be obtained by measuring end-tidal xenon solubility, and was proportional to both except in patients with severe lung disease causing significant respiratory dead zones. The entire data acquisition process takes 5-6 minutes, while the entire process, including data processing and CBF image generation, takes approximately 20 minutes. There is a 20-minute interval between each data acquisition, primarily to allow for xenon clearing.