Clinical application of intra-aortic balloon counterpulsation (IABP)

Intra-aortic balloon pump (IABP) is a widely used and effective treatment for left ventricular assist. The procedure involves inserting a catheter with a balloon into the arterial system, distal to the opening of the left subclavian artery within the descending aorta. This allows for cyclical filling and emptying of blood within the aorta, causing a phasic change in blood flow. This reduces left ventricular load, increases diastolic pressure, and enhances coronary perfusion.

I. Indications and Contraindications of IABP

1. Indications for IABP

(1) Pump failure can be caused by a variety of factors, including cardiogenic shock following acute myocardial infarction, perioperative myocardial infarction, low cardiac output syndrome after cardiopulmonary bypass, cardiac contusion, viral myocarditis, and toxic shock.

(2) Complications may occur after acute myocardial infarction, including ventricular septal perforation, mitral regurgitation, papillary muscle rupture, and ventricular aneurysm.

(3) For unstable angina that does not respond well to medical treatment.

(4) Refractory ventricular arrhythmias caused by ischemia.

(5) Expanded indications: Patients with left main coronary artery disease require special protection during interventional treatment, while high-risk patients or those whose interventional treatment has failed need more support. Sufficient support is also required during the perioperative period for cardiac surgeries such as coronary artery bypass grafting and valve replacement. Circulatory support is also essential before and after heart transplantation or artificial heart implantation in patients with end-stage heart disease. Comprehensive support is also required for high-risk heart disease patients undergoing major non-cardiac surgeries.

2. Contraindications for IABP

Severe aortic dissection, severe stenosis or calcification of the descending aorta or iliac artery, moderate to severe aortic regurgitation, irreversible brain damage, late-stage heart disease or other illnesses, and severe coagulation disorders are all contraindications.

II. Introduction to IABP Device

1. IABP Device

The counterpulsation machine consists of a main unit and a display screen. To use it, first connect the power cord, turn on the main power switch, and after the system self-test passes, open the helium cylinder valve to confirm the helium pressure. Connect the ECG leads and pressure sensor, and connect the pneumatic extension tubing. Press the start button to begin counterpulsation, adjusting the mode parameters as needed.

2. IABP Balloon Catheter

The IABP balloon catheter should not completely block the aortic lumen. Ideally, after inflation, the balloon should occupy 85%-90% of the aortic lumen, with a volume greater than 50% of the heart’s stroke volume. Complete blockage may cause damage to the aortic wall, as well as red blood cells and platelets. The IAB balloon volume is generally selected based on the patient’s height.

3. IABP Insertion

The intra-aortic balloon catheter is inserted through the femoral artery and into the descending aorta. Once the balloon catheter is in the descending aorta, its position must be confirmed by fluoroscopy or chest X-ray. The balloon should be positioned 1-2 cm below the left subclavian artery and above the renal artery (some literature suggests 2-3 cm). A daily chest X-ray should show the tip of the balloon between the second and third intercostal spaces. Correct positioning is crucial for the IABP to function effectively and to prevent blockage of important arteries such as the left subclavian artery and renal artery.

4. IABP Hemodynamics

The IABP catheter is inserted via a peripheral artery, and a balloon is placed in the descending aorta. During diastole, the balloon inflates to increase diastolic blood pressure, thereby increasing coronary artery (coronary artery) blood flow. In early systole, the balloon rapidly deflates to reduce cardiac afterload. The hemodynamic effects of IABP include improved coronary blood supply, reduced cardiac afterload and ventricular wall tension, a moderate increase in cardiac output (CO), and maintenance of blood pressure.

5. IABP Parameter Selection

(1) Trigger Modes: Four trigger modes are available: ECG trigger, pressure trigger, intrinsic frequency, and pacing mode. ECG trigger and pressure trigger are the most commonly used. ECG trigger uses the R wave as the trigger signal, while pressure trigger uses the rising systolic pressure wave as the trigger signal. Intrinsic mode inflates and deflates at a set frequency, suitable for situations where there is no ECG signal or pressure output. Pacing mode is suitable for patients with pacemakers.

(2) Counterpulsation Frequency: Generally, selecting 1:1 counterpulsation means counterpulsation occurs with every heartbeat. 1:2 or 1:3 counterpulsation means counterpulsation occurs only after 2 or 3 heartbeats, respectively.

(3) Inflation/Deflation Timing: Adjust the control key according to the ECG or pressure waveform to start inflating the balloon at the T-wave endpoint (triggered by ECG waveform) or the pressure notch (triggered by pressure waveform); and to start deflation before the QRS wave or at the aortic end-diastolic pressure point.

(4) Counterpulsation pressure: The counterpulsation pressure can be selected according to the patient’s hemodynamic status, that is, the inflation volume of the balloon can be adjusted.

III. Complications of IABP

1. Limb ischemia: The balloon catheter needs to be removed. If severe limb ischemia persists after removal, surgical treatment should be considered.

2. Bleeding and hematoma at the puncture site: Bleeding can be stopped by applying pressure to the puncture site, but good distal blood flow must be ensured. If bleeding cannot be stopped, surgical intervention should be considered.

3. Infection: The controllability of the infection and the necessity of removing the balloon catheter should be evaluated.

4. If balloon perforation occurs, the patient will experience the following symptoms: the counterpulsator will sound an alarm, blood spots will appear in the catheter, and the counterpulsation pressure waveform will suddenly change. If balloon perforation is suspected, counterpulsation therapy must be stopped immediately, the balloon catheter removed, and the patient placed in a supine position with their head down. If the patient still requires IABP treatment, a new balloon catheter will need to be inserted.

5. Thrombocytopenia: Platelet counts should be monitored dynamically, and platelet transfusions should be administered if necessary.

6. Aortic dissection: This may manifest as back or abdominal pain, decreased blood volume, or hemodynamic instability.

7. Thrombosis: The presentation and treatment of thrombosis should be determined based on the organ damaged. Strict anticoagulation is required throughout the IABP procedure.

IV. Indications for IABP withdrawal

1. Cardiac index > 2.5 L/min·m²;

2. Systolic blood pressure > 100 mmHg;

3. Maximum arterial pressure (MAP) > 80 mmHg;
4. Conscious, with good peripheral circulation, and urine output > 1 ml/kg·h;

5. No arrhythmias or myocardial ischemia on electrocardiogram;

6. Dopamine dosage or dobutamine < 5 μg/kg·min;

7. If the patient’s hemodynamics are stable under 1:4 ratio aortic balloon counterpulsation, this is an indication for removal of the intra-aortic balloon counterpulsation catheter.