Brain protection device
Brain protection device
Currently, both ex vivo and transcranial Doppler ultrasound examinations show that embolism is prone to occur during carotid artery stenting (CAS) when the balloon is inflated. With the increasing sophistication of brain protection devices (CPDs), the routine use of CPDs in carotid artery stenting has become a consensus among experts. In clinical treatment, any type of protective device can effectively block at least 60% of detached debris. However, the use of CPDs does not completely eliminate embolic complications; in fact, it may induce related complications such as difficulties in device insertion and release, problems with embolus capture effectiveness, vascular injury, and retrieval difficulties. Therefore, understanding and rationally using different types of CPDs is of great significance in reducing complications.
Classification and Common Products of Brain Protection Devices
CPDs are categorized into occlusive and non-occlusive types. Occlusive types include proximal and distal occlusive types; non-occlusive types include nickel-titanium internal support filters and non-supported air bag designs, each with its own advantages and disadvantages. Familiarity with the device design characteristics and its application to appropriate lesions and anatomical sites is crucial. Commonly used CPDs in my country include Angioguard, Filterwire, Spider (distal filter type), and MOMA (proximal occlusive type).
Use of remote filter-type brain protection devices
The advantage of distal filter-type CPD lies in its ability to maintain unobstructed blood flow throughout the procedure, thereby improving the safety of CAS (Cardiac Arthroplasty) procedures. The filter structure is placed on the guidewire itself (0.014in). After pre-shaping the flexible tip of the guidewire, it is passed through the stenotic lesion segment. If conditions permit, the procedure can be performed under “roadmap” guidance. This step, as well as subsequent steps such as filter release and retrieval, are performed without brain protection devices. Approximately 0-1% of patients may experience related neurological complications.
The filter should be placed in a relatively straight segment of the vessel 3-4 cm distal to the lesion (anchoring zone). Severe calcification of the vessel may make filter passage difficult. When the patient’s lesion is highly stenotic, pre-dilation can be performed using standard coronary artery manipulation techniques before filter placement, although this may carry risks. After filter release, internal carotid artery spasm can be avoided to prevent its movement within the vessel.
Unlike AngioGuard and FilterWire, Spider first passes the lesion segment with a 0.014in guidewire, and then advances the delivery system along the guidewire. Although the procedure is complex, it increases the selection of catheters and guidewires, improving the ability to pass through lesions. Different filter designs, due to differences in the materials of the guidewire, filter, and delivery system, have varying abilities to pass through lesions.
Use of proximal occlusive CPD
The design principle of this type of CPD is to block the proximal end of the lesion, causing backflow of blood in the internal carotid artery. After stent placement, residual blood and debris are actively aspirated and removed. The advantage is that blood flow is blocked before contact with the lesion, preventing stroke caused by CPD placement; however, patients must tolerate proximal blood flow occlusion. MOMA is a representative proximal occlusion CPD available in my country, integrating brain protection and embolism removal functions, with a 6F operating system. It uses two independent low-pressure compliant balloons to occlude the external carotid artery and common carotid artery; the maximum diameter of the common carotid artery balloon is 13mm, and the maximum diameter of the external carotid artery balloon is 7mm. Debris can be aspirated during or after the procedure.
MOMA (Modular Carotid Artery) Procedure:
- A 9F 25cm short sheath is placed via the common femoral artery; the MOMA is then inserted into the external carotid artery and common carotid artery via a 0.035in stiffened guidewire using coaxial guidewire exchange technology;
- After repeated angiography to confirm location and orientation, the distal balloon (external carotid artery) is inflated, followed by the proximal balloon (common carotid artery); contrast agent is manually advanced 2-3cm to confirm positive flow occlusion, and patient tolerance is observed;
- After flow occlusion, routine treatment of carotid artery lesions is performed; after stent deployment and re-dilation, blood is aspirated through the proximal side port of the MOMA system to remove embolic debris; after confirming the filter is free of debris, the balloon is released, and the MOMA system is withdrawn.
During the procedure, distal “reflux pressure” after occlusion needs to be monitored to help assess patient tolerance. If the reflux pressure remains stable above 30 mmHg, the patient’s tolerance is considered good.
Selection of brain protection devices
The three most commonly used distal filtering brain protection devices (CPDs) in China are AngioGuard, FilterWire, and Spider. They all share some common characteristics: after being placed distal to the stenotic segment of the carotid artery, they can effectively intercept fragmented atherosclerotic plaques that dislodge during balloon dilation and stenting, thereby preventing intraoperative cerebral embolism complications. The application of these CPDs can reduce the incidence of embolic complications in endovascular carotid artery angioplasty from 8% to 1%-2%.
In terms of operation, AngioGuard and FilterWire are relatively simple. Both use a protective cannula to deliver the protective umbrella through the stenotic artery segment to the distal end. The protective umbrella is connected to a 0.014-inch diameter delivery wire, available in 200cm and 300cm lengths, which can be selected according to the chosen stent type (standard overthe-wire, OTW, or rapid exchange).
The procedure for Spider-type CPDs is relatively complex: First, a 2.7F microprotective catheter is advanced through the stenotic segment of the carotid artery using a 0.014-inch guidewire. After withdrawing the 0.014-inch guidewire, the protective umbrella is then advanced through the catheter to the level of the skull base.
The protective umbrella retrieval process after stenting is largely the same for the three types of CPDs. A similar retrieval tube is used to insert and secure the protective umbrella, which is then safely withdrawn from the body. Because the three types of CPDs differ in length, texture, and structure, their ability to pass through lesions also varies. Therefore, the actual selection should be based on a comprehensive consideration of the lesion’s anatomical morphology, the degree of stenosis and calcification, and the operator’s experience. For lesions with severe stenosis, ulceration, tortuosity, or calcification that are difficult to pass through or prone to embolus dislodgement during passage, a proximal occlusion CPD may be considered.
Source:
Endovascular Surgery, edited by Guo Wei, People’s Military Medical Publishing House, November 2011.