cutting balloon
The cutting balloon (CB) has three or four sets of microblades attached to its surface. These blades are evenly distributed along the length of the balloon and around its circumference, with adjacent blades at 120° or 90° angles. When the balloon is folded, the blades are encased in folded wings. During balloon inflation, the blades create shallow incisions in the calcified plaque, causing the plaque to crack or rupture as expected at the incision site. This reduces intimal irregularities and severe tears, achieving dilation with relatively low pressure. Studies have shown that lumen loss 10 minutes to 24 hours after PTCA was significantly lower in the CB group than in the conventional balloon group (0.08–0.28 mm vs 0.20–0.33 mm, p=0.02). Compared to high-pressure balloons, CB can improve the primary patency rate of arteriovenous fistulas at 6 months in hemodialysis patients (relative risk 1.45; 95% confidence interval 1.08–1.96; p=0.01). However, the presence of the blade element also increases the outer diameter of the balloon when it is folded, reducing the product’s permeability within blood vessels.