Adjustable double balloon catheter
RenovoCath is an FDA-approved, innovative dual-balloon catheter drug delivery system developed by RenovoRx, serving as the core device of its transarterial microperfusion therapy platform.
The core mechanism of RenovoCath is based on pressure-mediated transvascular wall drug delivery. This technology uses dual balloons to create an isolation segment within the target vessel, establishing a pressure gradient between the vessel segment and the surrounding tumor tissue. This forces chemotherapy drugs to penetrate the arterial wall and directly enter the tumor tissue, rather than relying on traditional blood flow perfusion. This method is particularly suitable for poorly vascularized or poorly vascularized solid tumors (such as locally advanced pancreatic cancer), as these tumors often lack identifiable feeding arteries, making traditional transarterial chemoembolization (TACE) ineffective.
This technology platform utilizes vascular feeding vessels as the microscopic pathway for drug delivery, achieving targeted local administration and avoiding the toxic side effects of systemic intravenous chemotherapy. Clinical data show that this local delivery method significantly reduces systemic drug exposure while increasing local drug concentrations at the tumor site.
The RenovoCath catheter features a multi-lumen design, its core structure comprising a dual-balloon occlusion system, a multi-lumen body, and a dual-component handle. The catheter is equipped with two independently inflatable occlusion balloons, proximal and distal, controlled separately by a uniquely designed dual-component handle—a fixed handle controls the proximal balloon and keeps it stationary, while a movable handle controls the distal balloon, allowing physicians to adjust the distance between the two balloons (measured as interband distance using radiopaque markers) within a range of 15mm to 109mm. This adjustable design enables physicians to isolate vessel segments of varying lengths, eliminate collateral vessel interference, and create a closed drug perfusion environment.
The catheter’s internal structure is multi-lumen. The inner axis contains a distal balloon filling lumen and a guidewire lumen (compatible with 0.014-inch guidewires, extending to the distal tip); the outer axis is also multi-lumen, accommodating the inner axis while also providing a proximal balloon filling lumen and a drug infusion lumen (marked in red). Additionally, the catheter includes a flushing lumen for preoperative irrigation, forming a separate guidewire lumen (marked in blue) for easy guidewire manipulation. The catheter has an outer diameter of 0.076 inches and an effective working length of 75-85 cm (adjustable), compatible with 6F and 7F guiding catheters.
The balloon diameter is adjustable from 4-11 mm, inflated with a 70/30 saline-contrast mixture for precise positioning under fluoroscopic guidance. A radiopaque marker band between the two balloons identifies the total occlusion area, allowing for real-time image monitoring. This design forms a “closed-loop system,” ensuring that the infused fluid is confined within the isolation segment between the two balloons. Drug delivery is achieved by using a pressure gradient to drive the drug through the vessel wall into the tumor tissue, enabling targeted delivery.
This device is FDA approved for use in peripheral vascular systems for blood flow isolation, fluid delivery (including diagnostic or therapeutic drugs), and temporary vascular occlusion. It is suitable for applications such as arteriography, preoperative embolization, and chemotherapy drug infusion. Its unique advantage lies in overcoming the limitations of traditional single-balloon or fixed-balloon catheters in terms of flexibility, providing a precise control method for targeted therapy of visceral and peripheral vascular systems.